Compartir
Título
Influence of cycle length variations on antitachycardia pacing effectiveness among ICD patients
Autor(es)
Palabras clave
Antitachycardiapacing
Implantablecardioverter- defibrillator
Ventriculartachycardia
Fecha de publicación
2013-02
Editor
Elsevier
Citación
Jiménez-Candil, J., Hernández, J., Martín, A., Moríñigo, J., López, R., Ledesma, C., & Martín-Luengo, C. (2013). Influence of cycle length variations on antitachycardia pacing effectiveness among ICD patients. Heart Rhythm, 10(2), 207-213. https://doi.org/10.1016/J.HRTHM.2012.10.028
Resumen
[EN]Antitachycardia pacing (ATP) fails to terminate 5% to 25% of ventricular tachycardias (VTs) occurring in implantable cardioverter-defibrillator patients. We speculated that small fluctuations in VT cycle length (CL) may be related to the efficacy of subsequent ATP.
The purpose of this study was to determine the relationship between the R-R variations of the last 12 R-R intervals before ATP and the efficacy of the first ATP attempt.
We studied 551 VTs (CL 329±35 ms) occurring in 67 patients. We also analyzed the percentage of variation (P-RR), which was calculated by dividing the mean difference between each R-R interval and the next one by the CL (×100), and the acceleration index (AI), which was calculated by dividing the CL of the first 6 R-R intervals by the CL of the next 6.
The effectiveness of the first ATP therapy was 81%, being higher in VTs with AI<1 (85% vs 64%; P<.001). After classifying the events according to the tertiles of P-RR, ATP efficiency was better in higher values of P-RR (VTs with AI<1): 99% (third tertile) vs 85% (second tertile) vs 76% (first tertile), P<.001; and for VTs with AI≥1: 94% vs 68% vs 42% (P<.001). By logistic regression, P-RR (%; odds ratio 2.37; P<.001), and AI<1 (odds ratio 4.17; P<.001) were found to be independent predictors of successful first ATP attempts.
Small changes in CL increase the effectiveness of ATP significantly. VTs with lower degrees of R-R fluctuations, especially when the pattern is a progressive CL shortening, are infrequently terminated by ATP.
URI
ISSN
1547-5271
DOI
10.1016/j.hrthm.2012.10.028
Versión del editor
Aparece en las colecciones